    Once-off Service of BSC Cabinets and Filters 2026

    Tender: Once-off Service of BSC Cabinets and Filters

    • Tender Number: RFQ 670686
    • Department/Company: National Health Laboratory Service (NHLS) – Universitas, Bloemfontein
    • Closing Date: 30 March 2026 at 11:00
    • Location: Bloemfontein, Free State
    • Category: Laboratory Equipment, Reagents, Consumables, and Repair/Maintenance Services

     Tender Overview

    The NHLS Universitas Bloemfontein is seeking a qualified service provider for a one-time service of Biological Safety Cabinets (BSC) and filters.

    The objective is to ensure proper decontamination, certification, and maintenance of all BSC cabinets, maintaining laboratory safety, compliance with national/international standards, and operational integrity.

    Scope of Work

    The contractor will be responsible for:

    1. Decontamination:
      • Use safe and effective methods (e.g., hydrogen peroxide)
      • Ensure minimal disruption to laboratory operations
    2. Certification:
      • Verify cabinet integrity, HEPA filter leaks, airflow, negative pressure, smoke pattern, alarms, and interlocks
      • Perform testing according to EN12469 and NSF/ASI 49 standards
    3. Maintenance:
      • Replace HEPA filters only when required
      • Provide calibration and verification certificates for documentation
    4. Documentation:
      • Submit service reports, test results, and calibration certificates
      • Ensure traceability of equipment testing to national standards

     Technical Requirements

    • Decontamination using hydrogen peroxide with appropriate safety protocols
    • Certification performed by qualified technicians trained in EN12469 and NSF/ASI 49
    • Proper inspection and testing of airflow, containment, alarms, and interlocks
    • Replacement of HEPA filters only when necessary, matching specified models
    • Provision of calibration certificates for testing equipment

     Skills & Eligibility Requirements

    • Qualified technical staff with EN12469 and NSF/ASI 49 training/certification
    • Proven experience in decontamination and certification of BSC cabinets
    • Ability to provide accurate testing, calibration, and performance documentation
    • Submission of CVs for at least 3 personnel, including 2 SCM consultants and 1 project manager
    • References and proof of previous similar projects

     Submission Details

    • Submission Location:
      NHLS Bloemfontein Main Reception, RFQ Box, Cnr Hospital & De Korte Street
    • Reference Number on Quotation: 670686
    • Required Documents: As listed in the RFQ documentation
    • No briefing session mentioned

     Contact Information

    • For additional enquiries, download and review the full RFQ documents from the NHLS portal.

     